Helium Hydrogen Mass Spectrometer Leak Detector- High Sensitivity, Fast Response, Automatic Mobile Leak Detector For Industrial Use [KSNH-400]

1. Product Application and General Requirements

The KSNH-400 series leak detector is a fully automatic mobile helium mass spectrometer leak detector with a silent wheel design, suitable for use in many demanding industries. It adopts the principles of helium mass spectrometry and reverse diffusion, utilizes a 180° non-uniform magnetic field and fully automatic control technology, and realizes automatic helium peak scanning, automatic calibration and automatic range switching. It features short startup time, fast calculation speed, high detection sensitivity and strong anti-interference ability.

2. Performance Characteristics

  • Unique gas path design to prevent dust from entering the interior of the instrument.
  • Multi-mode function input and output for easy connection to various automation equipment.
  • Double iridium ribbon ion source with strong oxidation resistance and long service life.
  • Effective separation of machinery and circuits to avoid mutual interference from temperature, electromagnetic fields, etc., making the leak detector operate more stably and reliably.

3. Product Requirements and Main Specifications (Table)

Minimum detectable leak rate (Vacuum mode) 3×10⁻¹³ Pa·m³/s
Leak rate display range (Vacuum mode) 1×10⁻² Pa·m³/s ~ 1×10⁻¹³ Pa·m³/s
Startup time ≤ 120 S
Response time < 0.3 S
Maximum pressure at leak detection port 2000 Pa
Rough detection vacuum degree at leak detection port 2000 ~ 200 Pa
Medium detection vacuum degree at leak detection port 200 ~ 40 Pa
Fine detection vacuum degree at leak detection port Below 40 Pa
Helium pumping speed of the leak detector 5 L/s
Display units Pa·m³/s, mbar·l/s, atm·cc/s
Display interfaces Curve graph, histogram, numerical display
Leak detector dimensions 870×523×1166 (unit not specified, generally mm)
Control mode One-key fully automatic control; automatic zero adjustment; automatic calibration; automatic/manual switching of double filaments
Display mode LCD touch screen display (rotatable by 270°); configurable with 4 groups of alarm values
Core components and technology German imported turbo high-vacuum molecular pump; German BURKERT solenoid valve; imported AD chip; mechanical pump (Baosi); new mass spectrometry technology with 180° ionization deflection; equipped with standard leak hole and calibration certificate

4. Operating Environment

  1. This product is for indoor, industrial, and laboratory use only.
  2. Temperature: 5℃ ~ 40℃; Relative humidity: ≤ 80%.
  3. Operating voltage: Single-phase 220V ± 10%, 50Hz.
  4. Maximum operating current: 10A.